Use the hemostasis valves to impede the back-flow of blood during the implant procedure
and minimize the risk of air ingress.
Do not use a power injection syringe to inject contrast solution.
Precautions
This device should be used only by physicians who are trained in standard transcatheter
techniques. The physician should determine which patients are candidates for procedures
that use this device.
The physician should exercise clinical judgment in situations that involve the use of
anticoagulants or anti-platelet drugs before, during, and/or after the use of this delivery
catheter.
Use on or before the last day of the expiration month that is printed on the product
packaging label.
Use caution and rely on imaging guidance when advancing the sheath and dilator to avoid
damaging tissue and vessels or interfering with previously implanted medical devices.
Store in a dry place.
Adverse Events
Potential adverse events that may occur during or after a procedure using this device may
include, but are not limited to:
Air embolism
•Arrhythmia
Arteriovenous fistulae
Bleeding at the access site
Brachial plexus injury
Cardiac tamponade
•Death
Dissection
•Embolism
Endocarditis
Hematoma
Hemodynamic compromise
Infection
Myocardial infarction
Perforation
Peripheral pulse loss
Stroke/transient ischemic attack
Thrombosis
Valve damage
Vascular access site injury
Vascular occlusion
Vessel damage
Device Compatibility
Refer to the instructions for use supplied with the device to be implanted to determine catheter
compatibility.
Procedures
1. Prepare the delivery catheter for use:
- Inspect the delivery catheter sterile pouch and verify that it is unopened and
undamaged. Do not use the delivery catheter if the sterile barrier has been
compromised.
- Gently open the sterile pouch and inspect the components for damage. Do not use
damaged or kinked components.
- Flush all components with sterile saline.
- Wipe the delivery catheter with sterile gauze moistened with sterile saline to remove
any foreign material.
2. Connect the Tuohy-Borst hemostasis valve to the delivery catheter. Make sure there
are no gaps between the components.