DISINFECTING
i-STAT ALINITY INSTRUMENT
Disinfection is recommended between each patient. When instrument is dedicated
to a single patient, disinfect at least once a day. The disinfection process must begin
IMMEDIATELY
after the cleaning procedure is complete. Standard precautions
should be taken whenever working with blood or blood products.
Remove a wipe from the
container and squeeze out
any excess uid before you
begin to wipe the surface.
Wipe all surfaces three times.
This includes front, back, sides,
top and bottom.
WIPEWIPE
3X 3X
WIPE WIPE
3X 3X
W
I
P
E
3
X
3
X
1 2
3 4
Wipe with gauze
until dry.
Allow all
surfaces to
remain wet
for 3 minutes.
i-STAT Alinity Instrument
Cartridge Port
10-Pin Connector under the camera
Gold contacts (2) on the outside of
the battery
A
B
C
C
B
A
Avoid forcing liquid into these areas:
SENSITIVE AREAS