Table 6. Possible causes and soluons for potenal
issues
Problem Possible Cause Possible Soluon
the on-screen
nocaon to exit MRI
mode.
Program strength is
o.
The generator could
not deliver the desired
program strength.
Try adjusng the
program strength. If the
problem persists, contact
your clinician.
Increasing Strength
is Not
Recommended
message appears.
The program may need
to be modied.
Contact your clinician.
Bluetooth® Wireless Communicaon or Connecon
Paent controller
cannot establish
Bluetooth wireless
connecon with the
generator or the
connecon was lost.
Interference in the
environment (for
example, from
microwave ovens and
WiFi‡ routers).
Move to a dierent area
or room away from the
interference.
Paent controller is o
or has med out and is
in standby mode.
Wake up the paent
controller.
Paent controller
baery is drained.
Charge the paent
controller baery.
Bluetooth wireless
technology is not
connected.
Remove and re-enable
the Bluetooth wireless
technology pairing.
In the device sengs,
select Bluetooth. Tap the
toggle buon to enable
Bluetooth wireless
technology.