EasyManua.ls Logo

Abbott NeuroSphere 3875 - Quality of Service for Wireless Technology

Abbott NeuroSphere 3875
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The radio receiver in the device is using the same
frequency and bandwidth as the transmier.
Cables and transducers:
Cables and transducers are not used during normal use
of the device nor while programming the device.
Quality of Service for Wireless
Technology
Bluetooth® Low Energy (LE) wireless technology enables
communicaon between the generator and the paent
controller. The requirements for the quality of service
(QoS) vary depending on the use environment
(operang room, recovery room, and home
environment).
Aer the paent controller is paired with a generator,
the Bluetooth wireless technology symbol is visible on
the paent controller in the upper right-hand corner of
the screen. When the
Bluetooth wireless connecon is
not acve, the symbol is grayed out.
The standard Bluetooth LE wireless technology data rate
is 1 Mbit/s. The quality of service (QoS) should allow
wireless data to be transferred at a net rate of 2.5 kB/
sec. Each connecon interval includes a semi-duplex
transmission with a required acknowledge, a
transmission latency in each direcon (2x), and a
receive-to-transmit mode (RX-to-TX) me. Data is resent
82