13 |
SMART PIPETTE - USER GUIDE
SMA8-200
SMA12-200
20-200
µl
0.2
20 5 1.4
100 1 0.4
200 0.7 0.25
SMA8-300
SMA12-300
30-300
µl
0.2
30 3 1
150 1 0.5
300 0.6 0.3
Calibration Adjustment
1. Calibration adjustment is done with the Recalibration (Ellen) key.
2. Remove the calibration guard and place the Recalibration (Ellen)
key into the opening of the calibration screw on the side of the
handle.
3. Turn the key clockwise to increase or counterclockwise to de-
crease the volume.
4. After adjustment, check the calibration according to the instruc-
tions above.
5. Once within permissible error range, place calibration guard in its
position to avoid accidental changes in calibration.
For more detail information about Calibration and Adjustment, visit
www.accumaximum.com/calibration
Depending upon use, we recommend checking of calibration every six months.
However this can be adjusted to individual requirements.