11
Calibration
Note: The Calibration must be only accessed by official office of weights and
measures inspectors or licensed calibration technicians.
8.1 Destroy the Sealing
Upside down the scale and then find out the position of
sealing as the picture shows.
Destroy the sealing by tearing the fragile lead sealing. Then,
take off the sealing cover.
Use metal to make a short circuit on the two short pins lightly
as the picture shows.
After hearing a long ‘beep’ sound, the scale enters into Factory Mode, gravity acceleration
compensation value adjustment and zero/full weight learning can be carrying out.
Note: If the weight is not calibrated correctly after entering into Factory Mode, please do not long press
[C/ESC] key to exit Factory Mode!