SOPRO 717 First • User’s manual
22
9.1. HANDPIECE OR CONNECTION BOX MAINTENANCE
NOTE:
In case of contact with blood or excessive soiling, it is strongly recommended to follow a disinfecting process. First of all, clean the
handpiece with disinfecting wipes, then wrap the handpiece in several disinfecting wipes and leave for 15 minutes.
DESCRIPTION RECOMMENDATIONS
USE INSTRUCTIONS AND PRECAUTIONS

Disinfecting Surface cleaning and disinfecting
wipes e.g. Septol™ Wipes from
Pierre Rolland.
Take the wipe, remove excess
moisture, and then wipe
the equipment until visible
cleanliness is obtained.
Allow to dry in the open air.
Carefully close the packaging
box.
Do not scrub
Do not rinse.
Do not immerse in a
disinfecting liquid.