14
11. Disposal
The thermometer contains no hazardous materials. Remove the batteries
before disposal. Discard both thermometer and batteries in accordance with
national or local regulations.
12. Technical Specifications
Measurement unit: °F / °C
Operating mode: Adjusted mode (Body mode)
Direct mode (Surface mode)
Measuring site: Forehead
Reference body site: Axillary
Measurement range: Body mode: 93.2°F to 109.4°F (34.0°C to 43.0°C)
Surface mode: 32.0°F to 212.0°F (0°C to 100.0°C)
Accuracy: Body mode:
93.2°F to 94.8°F: ±0.5°F (34.0°C to 34.9°C:±0.3°C);
95.0°F to 107.6°F: ±0.4°F (35.0°C to 42.0°C:±0.2°C);
107.8°F to 109.4°F: ±0.5°F (42.1°C to 43.0°C:±0.3°C)
Surface mode: ±3.6°F (±2°C)
Display resolution: 0.1°F / 0.1°C
Three-color backlight: 95.9°F to 99.2°F (35.5°C to 37.3°C):
(Color alarm) Green (Normal temperature)
99.3°F to 100.4°F (37.4°C to 38.0°C):
Yellow (Slight fever)
100.5°F to 109.4°F (38.1°C to 43.0°C):
Red (High fever)
Note:
Surface mode is always green backlight.
In Body mode,93.2°F to 95.7°F (34.0°C-35.4°C) is green backlight
Facial indication: Happy Face: 95.9°F to 99.1°F (35.5°C to 37.3°C)
Sad Face: 99.3°F to 109.4°F (37.4°C to 43.0°C)
Auto power off time: ≤18 seconds
Measuring time: ≤2 seconds
Measuring distance: 0.04in - 5.9in (0.01cm - 15cm)
10. Calibration Testing
This thermometer is calibrated at the time of manufacture. If the thermometer
is operated in accordance with these instructions, periodic readjustment is
not required. ADC recommends checking calibration on an annual basis or
whenever clinical accuracy of the thermometer is in question.