Caution: When the ambient temperature of the thermometer changes too
much, such as moving the thermometer from one place of lower temperature
to another place of higher temperature, allow the thermometer to remain in a
room for 30 minutes where the temperature is between 59°F to 104°F
(15°C to 40°C).
Caution: Wherever possible, take measurements at the same area of the
forehead. Temperatures measured at different measurement sites
(e.g., on the temples, on opposite sides of the head) can vary considerably.
Caution: Head coverings that cover the forehead, perspiration, or hair
may affect the accuracy of measurements.
Caution: Avoid exposing the device to external heat sources or holding
the thermometer for an extended period of time as this can impact accuracy.
Do not point thermometer directly at known heat sources other than the patient.
Caution: If measuring human forehead temperature, select "body" mode;
for measuring objects, select "surface" mode.
Caution: Do not store or use near strong electrostatic or strong magnetic fields.
Caution: This product is intended for use as a screening tool in the home,
in small commercial or industrial settings (e.g., offices, factories, warehouses,
restaurants, retails shops, movie theaters) and in medical settings.
Caution: Avoid taking temperature measurements for at least 30 minutes
after physical activity, bathing, swimming, consuming food or beverages, or
spending time outdoors.
Caution If using for high-volume, rapid mass screenings, it is recommended
that the unit be allowed to rest for up to 30 seconds every 5 to 10
measurements so that the IR temperature sensor can reset and stabilize.