EasyManua.ls Logo

AED CellAED - Continuous Battery Diagnostic; Heart Rhythm Analysis; Defibrillation Waveform; Shelf Life

AED CellAED
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
There is an Infant Mode button that the user must press
if the patient is an infant weighing less than 10 kg (22 lbs);
this is a manual action taken when the CellAED®
audio (voice) prompt invites them to “For infants, press
infant button.” The user must press the infant button
a second time to confirm and enable Infant Mode.
The CellAED® will issue audio prompt warnings prior
to shock delivery. Shock delivery is automated by
the CellAED®.
If the patient is transferred to emergency medical
personnel, the medical personnel can remove the
CellAED® electrode pads and replace with their own
defibrillation equipment, if available. The CellAED®
must be removed before any alternative defibrillator
is applied. To deactivate the CellAED®, stick the gel
sides of the electrode pads together for at least five
seconds. Audio (voice) prompts will acknowledge
when the CellAED® has been deactivated. Dispose
of the CellAED® safely and immediately according
to local council regulations.
5.2.4 CONTINUOUS BATTERY DIAGNOSTIC
The CellAED® is constantly aware of the battery status
and provides visual feedback to indicate the state of
readiness.
A blinking GREEN LED (every 30 seconds) indicates
the CellAED® is fully operational and ready to use.
A blinking AMBER LED (every 30 seconds) indicates
that the battery is low and the CellAED® should be
replaced immediately.
No LED indicates that the battery has expired and
the CellAED® should be replaced immediately.
5.2.5 HEART RHYTHM ANALYSIS
The CellAED® evaluates the patient’s heart rhythm.
Refer to Section 10.4, page 57 (ECG Interpretation
and Performance) for further information.
5.2.6 DEFIBRILLATION WAVEFORM
The defibrillation shock, using patented CellAED®
biphasic waveform technology, is a fully-discharging
capacitator method for delivering an energy-
efficient biphasic defibrillation waveform. The
waveform is automatically adjusted to compensate
for patient impedance. The CellAED® does not
detect patient motion.
5.2.7 SHELF LIFE
The CellAED® can be used up to 13 months from
the date of manufacture. This date is printed on the
CellAED® as a Date of Manufacture and indicates
the Use-by Date for the CellAED®.
Maintaining the CellAED® in an ideal storage condition
(15°C – 35°C (59°F – 95°F)) is required to maximise
the performance of the battery as well as the
CellAED® expected shelf life (refer to Section 6.4,
page 28).

Table of Contents

Related product manuals