60
IMMUNITY TEST
IEC 60601-1-2:2014/EN 60601-
1-2:2015
TEST LEVEL
COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT
Radiated RF EM Fields IEC 61000-4-
3:2006/ A1:2007+A2:2010
10 V/m, 20 V/m 80MHz to 2.5GHz 10V/m Portable and mobile RF communications
equipment should be used no closer to any
part of the CellAED®, including cables, than
is necessary. The recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter is shown in
the following table.
Recommended separation distance
d = 1.20 x P, 80MHz to 800MHz
d = 2.30 x P, 800MHz to 2.5GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Interference may occur in the vicinity of equipment
marked with the following symbol above.
(Continued page 61)