AUDIO (VOICE) PROMPT INSTRUCTIONS
CALL FOR HELP. REMAIN CALM.
Call your local emergency services, or if possible, have a bystander call the local emergency services. Remain calm.
FOR ADULT, APPLY PADS ACROSS
CHEST. FOR INFANT, APPLY PADS
TO CHEST AND BACK.
Apply the electrode pads as instructed. Be prepared beforehand to place the electrode pads correctly. Take note of the difference
between patients weighing greater than 10 kg (22 lbs) (both pads applied to the chest) and infants weighing less than 10 kg
(22lbs) (one pad applied to the chest and one applied to the back) (refer to Figure 9 (page 37) for patients weighing greater than
10 kg (22lbs) and Figure 14 (page 42) for infants weighing less than 10 kg (22lbs)).
FOR INFANTS, PRESS INFANT
BUTTON
For infants weighing less than 10 kg (22 lbs) (typically less than 1 year old), press the infant button (Figure 5(d), page 34)
within 5 seconds of this audio (voice) prompt to provide the correct therapy. If the button is not pressed, the audio (voice)
prompt is repeated once. Failure to use the correct setting may result in harm to the patient and/or ineffective treatment.
PRESS INFANT BUTTON TO
CONFIRM INFANT MODE
For infants weighing less than 10 kg (22 lbs) (typically less than 1 year old), press the infant button (Figure 5(d), page 34)
within 5 seconds of this audio (voice) prompt to confirm and enable Infant Mode. Failure to use the correct setting may
result in harm to the patient and/or ineffective treatment.
INFANT MODE ENABLED
Infant Mode has been enabled. The CellAED® will automatically adjust its therapy. No operator or user action is required.
ANALYSING HEART RHYTHM
The CellAED® has begun automatically analysing the patient’s heart rhythm. Do not touch the patient during the analysis.
DO NOT TOUCH PATIENT
Remove all contact of yourself and anyone else touching any part of the patient. Failure to do so may result in ineffective
treatment and/or harm to the operator or patient.