MODEL 5600 REPAIR & CALIBRATION MANUAL CAT# 1431.01
8
REVISION 7: June 11, 1999
FILE LOCATION: E.F. Disk - Manuals - Cal. Man. - Model 5600.DOC
10/27/99
d) Fault Display Check:
•
Disconnect C1 or C2 the "FAULT" indicator must be lit.
e) Battery Charge Check:
•
Be sure the good voltage is selected (110 V or 220 V).
•
Connect the power cord to the supply and check that the green LED
"CHARGING" is on.
MEASURED
FUNCTION RANGE STANDARD VALUE TOLERANCE
VALUE POS NEG MIN MAX
µ
Ω
2000 µ
Ω
1000 µ
Ω
993 µ
Ω
1007 µ
Ω
m
Ω
20 m
Ω
1.000 m
Ω
.98 m
Ω
1.02 m
Ω
10 m
Ω
9.94 m
Ω
10.06 m
Ω
200 m
Ω
10 m
Ω
9.8 m
Ω
10.2 m
Ω
100 m
Ω
99.4 m
Ω
100.06 m
Ω
2000 m
Ω
100 m
Ω
98 m
Ω
102 m
Ω
1000 m
Ω
994 m
Ω
1006 m
Ω
Ω
20
Ω
1
Ω
0.98
Ω
1.02
Ω
10
Ω
9.94
Ω
10.06
Ω
200
Ω
10
Ω
9.8
Ω
10.2
Ω
100
Ω
99.4
Ω
100.6
Ω