EasyManua.ls Logo

AEMC 5600 - Control and Connector Identification

AEMC 5600
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODEL 5600 REPAIR & CALIBRATION MANUAL CAT# 1431.01
8
REVISION 7: June 11, 1999
FILE LOCATION: E.F. Disk - Manuals - Cal. Man. - Model 5600.DOC
10/27/99
d) Fault Display Check:
Disconnect C1 or C2 the "FAULT" indicator must be lit.
e) Battery Charge Check:
Be sure the good voltage is selected (110 V or 220 V).
Connect the power cord to the supply and check that the green LED
"CHARGING" is on.
MEASURED
FUNCTION RANGE STANDARD VALUE TOLERANCE
VALUE POS NEG MIN MAX
µ
2000 µ
1000 µ
993 µ
1007 µ
m
20 m
1.000 m
.98 m
1.02 m
10 m
9.94 m
10.06 m
200 m
10 m
9.8 m
10.2 m
100 m
99.4 m
100.06 m
2000 m
100 m
98 m
102 m
1000 m
994 m
1006 m
20
1
0.98
1.02
10
9.94
10.06
200
10
9.8
10.2
100
99.4
100.6

Other manuals for AEMC 5600

Related product manuals