MODEL 5600 REPAIR & CALIBRATION MANUAL CAT# 1431.01
8
REVISION 7: June 11, 1999
FILE LOCATION: E.F. Disk - Manuals - Cal. Man. - Model 5600.DOC
10/27/99
d) Fault Display Check:
Disconnect C1 or C2 the "FAULT" indicator must be lit.
e) Battery Charge Check:
Be sure the good voltage is selected (110 V or 220 V).
Connect the power cord to the supply and check that the green LED
"CHARGING" is on.
MEASURED
FUNCTION RANGE STANDARD VALUE TOLERANCE
VALUE POS NEG MIN MAX
µ
2000 µ
1000 µ
993 µ
1007 µ
m
20 m
1.000 m
.98 m
1.02 m
10 m
9.94 m
10.06 m
200 m
10 m
9.8 m
10.2 m
100 m
99.4 m
100.06 m
2000 m
100 m
98 m
102 m
1000 m
994 m
1006 m
20
1
0.98
1.02
10
9.94
10.06
200
10
9.8
10.2
100
99.4
100.6