Subject to Export Control, see Cover Page for details. 2-9
A. Duplex Test Screen
When the System Menu is displayed, press the 1 Key to access the Duplex Test Screen:
Edit Re tur n Hold Se tup PTT on
DUPLEX TEST
Generator Off Receiver More
MH
z: 40.000 000 MH z: 40. 000 00 0
Port: T/R Port: T/R
Level: -70 dBm Mod: FM 10k
Ext Attn dB: 0 AFBW: 0.3k HP
Modula tor Freq FM Analog
Ge n 1: On 10 00. 0 H
z 6.25 kH z
Ge n 2: On 15 0.0 H
z 3.25 kH z
MI C: On
6.25 kH z
Ext A ud: Off Load: High Z
Vo lu me: 1 00
Sq uelch: - 50 dBm
Sp eake r: Au dio In
AudOut : Audio In
Save Recall
41 20 29
(CTCSS)
Mo dulator Freq
Code FM DCS
Ge n 1: Off 15 0.0 H z 3.50 kH z
Dcs: Non
Inv 023 6 .25 kH z
MI C: Off 0 .00 kH
z
Ext A ud: Off Lo ad: High Z
Mo dulator
F
DTM
Ge n 1: Off 15 0.0 H z 3.50 kH z
DTMF: Burst Trig Lo: 6.25 kH
z
Seq: 0123 456789 Hi: 6.25 kH
z
Mark: 200 ms S pace : 200 ms
(DCS) (DTMF)
SCREEN FEATURE FUNCTION
Status (Generator) Displays the status (On / Off) of the Generator.
MHz (Generator) Used to select the signal generator frequency.
Range: 2.000000 to 1000.000000 MHz in 0.000001 MHz
increments
Port (Generator) Used to select the signal generator output connector.
Select: Ant, T/R or SWR
Receiver More
Ext A ttn dB: 0.0
Preamp: Auto