Aeroneb
®
Professional Nebulizer System Instruction Manual 29
Manual Cleaning
Cleaning nebulizer unit, T-pieces and neonate adapters
1. Ensure there is no medication remaining in the device.
2. Remove nebulizer unit from T-piece. Remove filler cap
from nebulizer unit.
3. Clean all parts with warm water and mild liquid detergent
in accordance with current hospital protocols.
4. Rinse parts with sterile water.
5. Shake excess water from parts and allow parts to fully
air dry.
CAUTION: Do not use abrasive or sharp tools to clean the
nebulizer unit.
Automated Washing Cycle
The Aeroneb Pro Nebulizer System has been validated with
the following automated washing cycles.
Automated Cycle One:
Detergent: Liquid alkaline cleaner (diluted as per
manufacturers instruction)
Water Quality: Mains water
Method:
1. Load the components in the automated washer.
2. Pre-rinse the components for 3 minutes.
3. Clean the components with liquid alkaline cleaner at 55°C
(131°F) for 10 minutes.
4. Rinse for 1 minute.
5. Rinse using thermal disinfection cycle at 93°C (199.4°F)
for 10 minutes.