30 Aeroneb
®
Professional Nebulizer System Instruction Manual
Automated Cycle Two:
Detergent: The following cycle was validated without the use
of a detergent
Water Quality: Mains water
Method:
1. Load the components in the automated washer.
2. Wash components for 10 minutes at 91°C (195.8°F).
3. Drain the machine for 40 seconds.
4. Rinse at 90°C (194°F) for 1 minute.
5. Drain the machine for 40 seconds.
6. Rinse at 90°C (194°F) for 1 minute.
7. Drain the machine for 40 seconds.
8. Dry at 90°C (194°F) for 15 minutes.
Disinfection of nebulizer unit, T-pieces and neonate
adapters
1. Follow steps 1 through 3 in manual cleaning section
on page 29.
2. Completely immerse parts in appropriate disinfecting
agent in accordance with current hospital protocols
and disinfectant agent manufacturer guidelines.
NOTE: Aerogen approves the following disinfection solutions
for use with its Aeroneb Pro Nebulization System regarding
material compatibility. With respect to microbiological
effectiveness, please ask the manufacturer. Refer to the
product labeling for specific instructions regarding activation,
safe use and disposal of these solutions.
Isopropyl (70%)
CIDEX
®
NU-CIDEX
®
CIDEX
®
OPA
Hexanios G+R