32 Aeroneb
®
Professional Nebulizer System Instruction Manual
NOTE: Sterilization using the long autoclave cycle ((iii)
above) may cause some areas of the nebulizer to become
discolored. This is not indicative of the performance of the
nebulizer unit.
b. To sterilize with hydrogen peroxide gas plasma,
place wrapped parts in a STERRAD
®
System, and
use the long cycle.
CAUTION: Users should refer to the product labelling for the
STERRAD
®
100S Sterilization System for specific instructions
regarding its proper operation.
Prior to next use:
(i) Check for cracks or damage and replace if any
defects are visible.
(ii) Perform a functional test as described in this
manual.
Cleaning of mounting brackets
Wipe clean with a damp cloth and mild liquid detergent. Do
not use abrasive or sharp tools.