Professional Nebulizer System Instruction Manual 33
Cleaning of control module, control module cable and
AC/DC adapter
1. Wipe clean with an alcohol based disinfectant wipe or a
quaternary ammonium compound based disinfectant
wipe.
2. Check for exposed wiring, damaged connectors, or other
defects and replace if any are visible.
3. Visually inspect for damage and replace the control
module if any damage is observed.
CAUTIONS:
Do not autoclave.
Do not use abrasive or sharp tools.
Do not spray liquid directly onto the control module.
Do not immerse control module in liquid.
Caution: The Aeroneb Pro nebulizer unit contains active
electronic components. Aerogen has validated the
methods of cleaning, disinfection and sterilization above.
The use of any other means of cleaning, disinfection or
sterilization has not been validated and is likely to reduce
the life of your nebulizer and will invalidate your warranty.