Professional Nebulizer System Instruction Manual 35
Table 3: Aeroneb Pro troubleshooting
If this happens:
It could mean:
Try this:
The 15 Min. or
30 Min. indicator
lights, but
aerosol is not
visible.
No medication in
nebulizer unit.
Refill medication
through filler cap in
the nebulizer unit
(see Adding
medication during
nebulization).
Nebulizer unit has
not been properly
cleaned.
Clean nebulizer unit
(see Cleaning and
sterilization).
It may be time to
replace the
nebulizer unit.
See Warranty and
Life of Product.
Refer to Aeroneb
Pro parts list.
15 Min. or
30 Min. indicator
does not light
when On/Off
power button is
pressed.
There is no power
to the System.
Verify that AC/DC
adapter is securely
attached to control
module.
Rechargeable
battery is
depleted.
Recharge battery
(see Recharging
the battery).
The fault
indicator lights.
The control
module cable is
incorrectly
connected, or
electronics are
malfunctioning.
Verify that control
module cable is
correctly connected
to both the
nebulizer unit and
the control module.