EasyManua.ls Logo

Aerogen Aeroneb Pro - Page 42

Aerogen Aeroneb Pro
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36 Aeroneb
®
Professional Nebulizer System Instruction Manual
Table 3: Aeroneb Pro troubleshooting
If this happens:
It could mean:
Try this:
Longer than
expected
treatment time.
e.g. 3 ml of
Normal Saline
(0.9%) should
take no longer
than 15 minutes to
nebulize
Rechargeable
battery is depleted.
Recharge battery
(see Recharging the
battery).
Nebulizer unit has
not been properly
cleaned.
Clean nebulizer
(see Cleaning and
sterilization).
It may be time to
replace the
nebulizer unit.
See Warranty and
Life of product.
Refer to Aeroneb Pro
parts list.
Medication is left
in the nebulizer
unit after
nebulization cycle.
Nebulizer was not
turned on or
connected to power.
Ensure that nebulizer
is connected to
power and turned on.
Rechargeable
battery is depleted.
Recharge battery
(see Recharging the
battery).
Nebulizer unit has
not been properly
cleaned.
Clean nebulizer
(see Cleaning and
sterilization).
A 15 minute cycle
was selected and a
volume greater
than 3 mL was
added to the
nebulizer unit.
Run an additional
15 minute cycle.
When delivering a
dose greater than
3 mL select the
30 minute cycle.
It may be time to
replace the
nebulizer unit.
See Warranty and
Life of Product.
Refer to Aeroneb Pro
parts list.
Note: The rechargeable battery in the control module should
only be replaced by Aerogen authorized personnel: contact
your Aeroneb Pro nebulizer system sales representative.

Table of Contents