Delete UserGroup Memory
The values stored in the memory are deleted by user ID.
You cannot partially delete values stored in the memory. All values for the
user you select will be deleted.
Select your ID (lor
2).
Pressthe M button first to view mode.
PressS button until" DEL" is displayed at the right corner of the screen,
then PressM button again all the values will be deleted.
Delete All Memory
Press" 8" button while the unit is turned off, then press three button
(8
+
0)
+@)
simultaneously, release all button after the screen flicker
all values will be deleted.
Abnormal phenomenon
Cause
Faulttreatment method
Position of arm cuff is not
Keep correct posture
kept consistent with height
during measurement.
of heart.
Moving or talking during
Keep silent and do not
Measured value istoo
measurement.
talk during measurement.
high or too low.
Arm and shoulder muscles
Stay relaxed and
are tensed.
measure again.
Arm cuff is too loose or
Wrap the arm cuff
too tight.
correctly.
Blood pressure value
Blood pressure value will chonge at
measured each time is
different measuring times, and with
Measuring posture is wrong.
different postures and degrees of
different (reading too low
reicxonon. Toke a deep breath
or too high).
before measuring 10 keep your
body relaxed
Shutdown during
Low battery.
Charge the product.
measurement.
No response when
Low battery.
Charge the product.
starting.
LCD Trror Tip and Corresponding Cause(table 1)
LCDerror tip
Couse
"Er 1"
Make sure the gas in the arm cuff was out; Don't press
the belt when measuring.
"Er 2"
Don't move when measuring.
"Er 3"
Abnormal test result. Please test again.
"Er 4"
Belt was loose or air leak.
"Er 5"
Trachea was stuck.
"Er 6"
Too big interference.
"Er 7"
Pressure beyond limit.
Blood Pressure Warning Strip Illustration (table
2)
LCD
High pressure
Low pressure
Severe
;'180
;'110
Moderate
160-179
100-109
Mild
140-159 90-99
Normal or High
130-139
85-89
Normal
120-129
80-84
Best