AFT-2000W User Manual
4.33 How to solidify the parameters
There are three kinds of parameters for the new generation TUF/TUC-2000:
1)
Current parameters, the parameters are stored in the RAM. They will be lost when one cut the
power or remove the reserve battery.
2)
Solidified parameters, you may use menu window 26 to store the parameters in the Flash
memory and they will not be lost even power off. This menu window is also a switch for the
parameters in flash memory to be loaded when power is turned on. The default option is that the
parameters will be loaded. So if the parameters are very stable, you may need this option.
3)
User frequently-used parameters, you can use menu window 27 to store or restore from the
internal Flash memory, as many as 9 different pipe parameter configurations.
4.34 How to enter the parameters of user-type-transducer
If a user-type-transducer is selected in menu window 23, you need proceed to enter additional 4
user-type-wedge parameters that describe the user transducers.
If the PI-type transducer is selected, you need enter additional 4 PI-type transducer parameters that
describe the PI-type transducers.
4.35 How to use the circular display function
When entering menu window 95, the circular display function will be started automatically. The
following windows will be displayed one by one, each window will stay for 8 seconds:
M95>>M00>>M01>>M02>>M02>>
M03>>M04>>M05>>M06>>M07>>M08>>M90>>M91>>M92>> M93>> M94>>M95. This
function allows the user to visit all the important information without any manual action.
To stop this function, simply press a key. Or switch to a window other than M95.
4.36 How to enter into the linearity correcting? How to enter into the
data?
When the product leaved the factory, the function is closed.
The menu can realize almost 12 parts linearity correcting. The user can choose from two points to
twelve points to execute the linearity correcting according to user actual condition.
In order to explain the usage method of the menu, we suppose that we get the following table data
through calibration the meter.
Reference standard
equipment flow (m
3
/h)
Instrumented show
flow (m
3
/h)
Correction coefficient
(standard / show value)
1.02
0.0998
1.02
5.11
5.505
0.93
10.34
10.85
0.95
20.45
19.78
1.03