4450A EN 20060221 Introducing the CR 85-X
CR 85-X DIGITIZER
Resetting the CR 85-X
In exceptional circumstances you may be prompted to reset the CR 85-X,
either by a message on the keypad or as part of a troubleshooting procedure in
this manual.
To reset the Digitizer:
1 Locate the main switch and place it in position ‘OFF’.
2 Wait 30 seconds.
3 Place the main switch in position ‘ON’.
Caution: Never reset the Digitizer to solve a plate or cassette jam. If you
would do so, the plate inside the Digitizer might get damaged. In case of a
plate or cassette jam, always follow the procedures described in Chapter 4,
‘Preventive maintenance and troubleshooting’ of the Reference manual.