31
4450A EN 20060221 Basic operation (‘Operator mode’)
CR 85-X DIGITIZER
Reading an emergency image plate
You may have an image plate which you wish to give priority over other image
plates which are being processed by the image processing station. Such image
plates are referred to as 'emergency image plates'.
To read an emergency image plate:
1 Check that the CR 85-X is ready for operation:
the CR 85-X must display the operator screen with ‘Ready’ or ‘Busy’ status.
the status indicator at the top of the CR 85-X must be constant or flashing
green.
Note: The emergency status will only be assigned to the first image plate
which you insert into the CR 85-X cassette slot after pressing the Emergency
key.
Note: The CR 85-X is operational if the status field equals ‘READY’, even if
status messages of the destination are shown (e.g. ‘VIPS not ready’).
VIPS not ready READY