EasyManua.ls Logo

AGFA DRYSTAR 5302 5366 - Safety Check Upon Completion of the Repair Work; Visual Check

AGFA DRYSTAR 5302 5366
465 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DD+DIS150.05E
Repair and Service
Machine specific safety instructions and hints for repair work
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0 DRYSTAR 5302 Chapter 3.1 / 5
2005-09-09 (Type 5366/100) Agfa Company Confidential
2 Safety check upon completion of the repair work
Upon completion of any repair job, especially after replacing or installing
conductive parts, check the following points as a routine safety check.
Country specific regulations must be observed as well.
Safety check Details see..
Visual check 2.1
Electrical check 2.2
2.1 Visual check
• Make sure that the power cord and the cables inside the machine do not
show any signs of damage.
• Make sure that the ventilation slots before the print head and the power
supply are free and that the print head and power supply ventilation is
working properly.
• Make sure, that no obstacles can cause stumbling (e.g. cables), and no
defective covers, sharp edges or parts sticking out in the vicinity of the
machine can cause injuries.
• Make sure that no smoldering spots or wrong cabling is visible.
• Make sure that all grounding connections to metallic covers are present.

Table of Contents