Table 8-2 (Cont.)
Ceiling Suspension Motor Controller Boards
CODE USER ERROR MESSAGE ERROR DESCRIPTIO N ACTION
11121 Board Failure: CS-Angular
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Change board
11122 End-Switch Limit: CS-Angular
Axis moved out of
defined range manually .
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range.
11123
Operative rejected: CS-Rotational
position out of calibrated range:
Right there, only MANUAL
movement allowed
Axis moved out of
defined range manually .
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check position feedback.
Contact Service
11124 Board Failure: CS-Rotational
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Change board
11125 End-Switch Limit: CS-Rotational
Axis moved out of
defined range manually .
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check End Switch Limit position,
activation, operation
Check End Switch Limit range
11126 Control Board Error: CS-Interface Change board