Table 8-2 (Cont.)
Ceiling Suspension Motor Controller Boards
CODE USER ERROR MESSAGE ERROR DESCRIPTIO N ACTION
11113
Operative rejected: CS-Transversal
position out of calibrated range:
Right there, only MANUAL
movement allowed
Axis moved out of
defined range manually .
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check position feedback
Contact Service
11114 Board failure: CS-Transversal
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Change board
11116
End-Switch Limit:
CS-Transversal
Axis moved out of
defined range manually .
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range
11117
Operative rejected: CS-Vertical
position out of calibrated range:
Right there, only MANUAL
movement allowed
Axis moved out of
defined range manually .
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check position feedback.
Contact Service
11118 Board Failure: CS-Vertical
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Change board
11119 End-Switch Limit: CS-Vertical
Axis moved out of
defined range manually .
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range
11120
Operative rejected: CS-Angular
position out of calibrated range:
Right there, only MANUAL
movement allowed
Axis moved out of
defined range manually .
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check position feedback
Contact Service