DX- D 600
User Manual
0242F EN 20181218
156
Table 8-8
Auto-centering
CODE USER ERROR MESSAGE ERROR DESCRIPTIO N ACTION
11801 Auto-movement aborted: CS moved
Auto-centering is cancelled because
Ceiling Suspension was moved
Try again
11802
Auto-movement aborted: Wall Stand
moved
Auto-centering is cancelled because
Wall Stand was moved
Try again
11803 Auto-movement aborted: Table moved
Auto-centering is cancelled because
Table was moved
Try again
11804
Auto-movement aborted: Detector
changed
Auto-centering is cancelled because
current Bucky/Detector was changed
Try again
11815
Auto-movement aborted: No Detector
selected
Auto-centering does not start because
NO Bucky/Detector is selected
Select Bucky/Detector and try again
11821
Operative rejected: CS-Longitudinal
Target is out of calibrated range:
Please, check POSITION parameters
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) is out of range on
Longitudinal Axis
Move Bucky/Detector and try again
11822
Auto-movement aborted:
The CS-Longitudinal Axis is blocked:
Please, try again
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) Longitudinal Axis
timed out
Try again
11831
Operative rejected: CS-Transversal
Target is out of calibrated range:
Please, check POSITION parameters
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) is out of range on
Transversal Axis
Move Bucky/Detector and try again
11832
Auto-movement aborted:
The CS-Transversal Axis is blocked:
Please, try again
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) Transversal Axis
timed out
Try again
11841
Operative rejected: CS-Vertical Target
is out of calibrated range: Please,
check POSITION parameters
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) is out of range on
Vertical Axis
Move Bucky/Detector and try again
11842
Auto-movement aborted:
The CS-Vertical Axis is blocked:
Please, try again
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) Vertical Axis timed
out
Try again