DX- D 600
User Manual
0242F EN 20181218
158
Table 8-9
Auto-tracking
CODE USER ERROR MESSAGE ERROR DESCRIPTIO N ACTION
11903
Tracking aborted: No Detector
selected
No Bucky/Detector selected Select Bucky/Detector and try again
11906
Operative rejected:
Detector and Tube-crane are
misaligned: Please, get a valid SID
Bucky/Detector and X-ray Tube are
misaligned
Please get the X-ray Tube aligned
11907
System geometry: Positioner out of
limits
Positioner out of limits
Please Place SID: between 80 and
300 cm
11913
Auto-tracking aborted:
Detector changed
Target Bucky/Detector changed
position
Try again
11914 Auto-tracking aborted: CS moved
Ceiling Suspension moved manually
and misaligned
Activate T racking function
11915
Auto-tracking aborted:
Wall Stand moved
Wall Stand moved manually and
misaligned
Activate T racking function
11916 Auto-tracking aborted: Table moved Table moved manually and misaligned Activate Tracking function
11921
Auto-tracking aborted:
CS-Longitudinal limit
Target position out of calibrated limits
on Longitudinal Axis of the Ceiling
Suspension
Calibrate the Ceiling Suspension
longitudinal Axis
11922
Auto-tracking paused:
CS-Longitudinal limit
Target position out of limits on
Longitudinal Axis of the Ceiling
Suspension
Do not exceed the longitudinal limit
11931
Auto-tracking aborted:
CS-Transversal limit
Target position out of calibrated limits
on Transversal Axis of the Ceiling
Suspension
Calibrate the Ceiling Suspension
transversal Axis
11932
Auto-tracking paused:
CS-Transversal limit
Target position out of limits on
Transversal Axis of the Ceiling
suspension
Do not exceed the transversal limit
11941
Auto-tracking aborted:
CS-Vertical limit
Target position out of calibrated limits
on Vertical Axis of the Ceiling
Suspension
Calibrate the Ceiling Suspension
vertical Axis
11942
Auto-tracking paused:
CS-Vertical limit
Target position out of limits on Vertical
Axis of the Ceiling suspension
Do not exceed the vertical limit
11981
Auto-tracking aborted:
Wall Stand-Vertical limit
Target position out of limits on Vertical
Axis of the W all Stand
Calibrate the Ceiling Suspension
vertical Axis
11991
Auto-tracking aborted:
Table-Longitudinal limit
Target position out of calibrated limits
on Longitudinal Axis of the Table
Calibrate the Table
11992
Auto-tracking paused:
Table-Longitudinal limit
Target position out of limits on
Longitudinal Axis of the Table
Do not exceed the Table limits