GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
(IEC 60601-1-2:2014)
This X-ray System is intended for use in the electromagnetic environment specified below.
The customer or Operator of this X- ray System should assure that it is used in such an environment.
Immunity Test
IEC 60601- 1-2:2014
Test Level
Compliance Level
Electromagnetic
environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
¦ 8 kV contact
¦ 2kV,¦ kV, ¦ 8kV,
¦ 15 kV air
¦ 8 kV contact
¦ 2kV,¦ kV, ¦ 8kV,
¦ 15 kV air
Floors should be wood, concrete or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity
should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
¦ 2 kV for power supply lines
¦ 1 kV for input/output lines
(100 kHz repetition frequency)
¦ 2 kV for power supply lines
¦ 1 kV for input/output lines
(100 kHz repetition frequency)
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
¦ 0.5 kV, ¦ 1kV
line(s) to line(s)
¦ 0.5 kV, ¦ 1kV,¦ 2kV
line(s) to earth
¦ 0.5 kV, ¦ 1kV
line(s) to line(s)
¦ 0.5 kV, ¦ 1kV,¦ 2kV
line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines.
IEC 61000-4-11
0% U
T
for 0.5 cycle
at 0
o
,45
o
,90
o
,135
o
, 180
o
,
225
o
, 270
o
and 315
o
0%U
T
for 1 cycle
at 0
o
70 % U
T
for 25/30 cycles
at 0
o
0% U
T
250/300 cycles
0% U
T
for 0.5 cycle
at 0
o
,45
o
,90
o
,135
o
, 180
o
,
225
o
, 270
o
and 315
o
0%U
T
for 1 cycle
at 0
o
70 % U
T
for 25/30 cycles
at 0
o
0% U
T
250/300 cycles
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the This
X-ray System requires continued
operation during power mains
interruptions, it is recommended
that this X-ray System is powered
from an Uninterruptible Power
Supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
NOTE - U
T
is the a.c. mains voltage prior to application of the test level.