Quantitative Information
Loading Factors Parameter Selection Filtrat. Measured Doses
KVp
mA
Time (s)
mAs
Focal Spot
Selection
SID
Source--Image Distance
(cm)
Collimator blades
opening
(cm)
HVL
(min. value allowed)
(mmAl)
Collimator
Output Dose
(μGy*m2)
SID Dose
(mGy)
Phantom
Input Dose
(mGy)
Phantom
Input Dose Rate
(Gy/h)
Phantom
Outp ut Dose Rate
(mGy/h)
Phantom
Output Dose
(μGy)
160 0.012 2 Small 0.2 0.016 0.025 7.479 10.795 0.036
100 0.1 10 Small
1.1 0.087 0.136 4.906 7.682 0.213
200 0.5 100 Large
13x13
11 0.836 1.307 9.407 14.125 1.962
400 1 400 Large
40 3.073 4.802 17.286 23.863 6.629
160 0.012 2 Small
100
2.1 0.016 0.025 7.615 18.691 0.062
100 0.1 10 Small
11.8 0.090 0.140 5.038 13.354 0.371
200 0.5 100 Large
43x43
107.1 0.862 1.347 9.698 23.798 3.305
400 1 400 Large
391.3 3.166 4.947 17.809 41.228 11.452
40
160 0.012 2 Small
1.6
0.2 0.005 0.006 1.865 4.273 0.014
100 0.1 10 Small
1.1 0.027 0.034 1.214 3.453 0.096
200 0.5 100 Large
13x13
11 0.257 0.325 2.343 5.985 0.831
400 1 400 Large
40 0.940 1.190 4.283 11.723 3.257
160 0.012 2 Small
180
2.1 0.005 0.007 1.962 6.243 0.021
100 0.1 10 Small
11.8 0.028 0.035 1.269 4.420 0.123
200 0.5 100 Large
43x43
107.1 0.267 0.338 2.432 7.400 1.028
400 1 400 Large 391.3 0.979 1.239 4.461 12.763 3.545
Combined standard uncertainty is
¦
35%
(IEC 60580:2000 / 60601-2-54:2009
and IEC 60601--2--54:2009/AMD1:2015).
Note .