DX- D 600
User Manual
0242F EN 20181218
47
Quantitative Information
Loading Factors Parameter Selection Filtrat. Measured Doses
KVp
mA
Time (s)
mAs
Focal Spot
Selection
SID
Source--Image Distance
(cm)
Collimator blades
opening
(cm)
HVL
(min. value allowed)
(mmAl)
Collimator
Output Dose
(μGy*m2)
SID Dose
(mGy)
Phantom
Input Dose
(mGy)
Phantom
Input Dose Rate
(Gy/h)
Phantom
Outp ut Dose Rate
(mGy/h)
Phantom
Output Dose
(μGy)
160 0.012 2 Small 0.6 0.046 0.072 21.746 113.713 0.379
100 0.1 10 Small
3.9 0.252 0.394 14.195 79.388 2.205
200 0.5 100 Large
13x13
39.4 2.587 4.042 29.103 157.649 21.896
400 1 400 Large
191.4 10.009 15.639 56.299 295.137 81.983
160 0.012 2 Small
100
7.5 0.048 0.074 22.299 233.322 0.778
100 0.1 10 Small
40.6 0.265 0.414 14.894 161.562 4.488
200 0.5 100 Large
43x43
389.3 2.691 4.205 30.277 320.682 44.539
400 1 400 Large
1491.3 10.435 16.304 58.696 596.348 165.652
60
160 0.012 2 Small
2.2
0.6 0.014 0.018 5.345 53.374 0.178
100 0.1 10 Small
3.9 0.078 0.098 3.538 36.438 1.012
200 0.5 100 Large
13x13
39.4 0.796 1.007 7.251 72.125 10.017
400 1 400 Large
191.4 3.078 3.896 14.025 145.377 40.383
160 0.012 2 Small
180
7.5 0.015 0.019 5.677 71.217 0.237
100 0.1 10 Small
40.6 0.082 0.103 3.717 48.584 1.350
200 0.5 100 Large
43x43
389.3 0.832 1.053 7.582 96.355 13.383
400 1 400 Large 1491.4 3.219 4.074 14.667 179.186 49.774
Combined standard uncertainty is
¦
35%
(IEC 60580:2000 / 60601-2-54:2009
and IEC 60601--2--54:2009/AMD1:2015).
Note .