DX- D 600
User Manual
0242F EN 20181218
48
Quantitative Information
Loading Factors Parameter Selection Filtrat. Measured Doses
KVp
mA
Time (s)
mAs
Focal Spot
Selection
SID
Source--Image Distance
(cm)
Collimator blades
opening
(cm)
HVL
(min. value allowed)
(mmAl)
Collimator
Output Dose
(μGy*m2)
SID Dose
(mGy)
Phantom
Input Dose
(mGy)
Phantom
Input Dose Rate
(Gy/h)
Phantom
Outp ut Dose Rate
(mGy/h)
Phantom
Output Dose
(μGy)
160 0.012 2 Small 1.4 0.087 0.136 40.753 378.000 1.260
100 0.1 10 Small
7.4 0.461 0.702 25.909 256.070 7.113
200 0.5 100 Large
13x13
74.5 4.674 7.303 52.582 511.763 71.078
400 1 400 Large
366.7 18.374 28.709 103.353 982.017 272.783
160 0.012 2 Small
100
14.3 0.090 0.141 42.391 829.043 2.763
100 0.1 10 Small
77 0.483 0.754 27.162 553.148 15.365
200 0.5 100 Large
43x43
735.9 4.884 7.632 54.949 1099.409 152.696
400 1 400 Large
2856.2 19.209 30.014 108.049 2111.165 586.435
80
160 0.012 2 Small
2.9
1.4 0.026 0.033 9.931 181.096 0.604
100 0.1 10 Small
7.2 0.142 0.179 6.462 120.177 3.338
200 0.5 100 Large
13x13
74.5 1.449 1.834 13.201 239.228 33.226
400 1 400 Large
366.7 5.703 7.218 25.986 480.835 133.565
160 0.012 2 Small
180
14.3 0.027 0.035 10.419 249.574 0.832
100 0.1 10 Small
77 0.149 0.189 6.799 162.094 4.503
200 0.5 100 Large
43x43
735.9 1.520 1.924 13.851 328.883 45.678
400 1 400 Large 2856.2 5.988 7.578 27.282 632.661 175.739
Combined standard uncertainty is
¦
35%
(IEC 60580:2000 / 60601-2-54:2009
and IEC 60601--2--54:2009/AMD1:2015).
Note .