Quantitative Information
Loading Factors Parameter Selection Filtrat. Measured Doses
KVp
mA
Time (s)
mAs
Focal Spot
Selection
SID
Source--Image Distance
(cm)
Collimator blades
opening
(cm)
HVL
(min. value allowed)
(mmAl)
Collimator
Output Dose
(μGy*m2)
SID Dose
(mGy)
Phantom
Input Dose
(mGy)
Phantom
Input Dose Rate
(Gy/h)
Phantom
Outp ut Dose Rate
(mGy/h)
Phantom
Output Dose
(μGy)
160 0.012 2 Small 2.1 0.131 0.205 61.550 854.348 2.848
100 0.1 10 Large
11.2 0.698 1.091 39.282 562.852 15.635
200 0.5 100 Large
13x13
113 7.136 11.149 80.276 1 132.591 157.304
400 1 400 Large
448.9 28.400 44.375 127.800 1784.097 619.478
160 0.012 2 Small
100
21 0.137 0.215 64.362 1829.478 6.098
100 0.1 10 Large
114.8 0.735 0.140 41.371 1221.809 33.939
200 0.5 100 Large
43x43
1067.6 7.491 1.347 84.277 2346.574 325.913
400 1 400 Large
4373 29.791 4.947 134.061 3901.774 1354.78
100
160 0.012 2 Small
3.6
2.1 0.040 0.006 15.334 396.261 1.321
100 0.1 10 Large
11.2 0.217 0.034 9.877 263.614 7.323
200 0.5 100 Large
13x13
113 2.224 0.325 20.269 536.807 74.557
400 1 400 Large
448.9 8.878 1.190 32.361 861.997 299.304
160 0.012 2 Small
180
21 0.043 0.007 16.187 555.391 1.851
100 0.1 10 Large
114.8 0.228 0.035 10.404 363.757 10.104
200 0.5 100 Large
43x43
1067.6 2.334 0.338 21.268 743.791 103.304
400 1 400 Large 4373 9.313 1.239 33.946 1173.788 407.565
Combined standard uncertainty is
¦
35%
(IEC 60580:2000 / 60601-2-54:2009
and IEC 60601--2--54:2009/AMD1:2015).