Quantitative Information
Loading Factors Parameter Selection Filtrat. Measured Doses
KVp
mA
Time (s)
mAs
Focal Spot
Selection
SID
Source--Image Distance
(cm)
Collimator blades
opening
(cm)
HVL
(min. value allowed)
(mmAl)
Collimator
Output Dose
(μGy*m2)
SID Dose
(mGy)
Phantom
Input Dose
(mGy)
Phantom
Input Dose Rate
(Gy/h)
Phantom
Outp ut Dose Rate
(mGy/h)
Phantom
Output Dose
(μGy)
160 0.012 2 Small 2.9 0.194 0.303 90.897 1611.652 5.372
100 0.1 10 Large
19.1 1.037 1.620 58.304 7.682 0.213
200 0.5 100 Large
13x13
164.1 10.722 16.753 120.620 2195.061 304.870
400 1 400 Large
823.7 43.078 67.310 121.158 2211.652 1228.696
160 0.012 2 Small
100
29.7 0.204 0.319 95.666 3558.261 11.861
100 0.1 10 Large
163.4 1.090 1.704 61.337 2407.617 66.878
200 0.5 100 Large
43x43
1595.2 11.243 17.568 126.489 4963.617 689.391
400 1 400 Large
5679.6 45.270 70.734 127.321 4418.609 2454.783
125
160 0.012 2 Small
4.5
2.9 0.058 0.073 21.923 776.609 2.589
100 0.1 10 Large
19.1 0.317 0.401 14.449 520.278 14.452
200 0.5 100 Large
13x13
164.1 3.349 4.238 30.515 1068.730 148.435
400 1 400 Large
823.7 13.470 17.047 30.685 1072.487 595.826
160 0.012 2 Small
180
29.7 0.062 0.078 23.395 1085.478 3.618
100 0.1 10 Large
163.4 0.338 0.428 15.416 728.765 20.243
200 0.5 100 Large
43x43
1595.2 3.523 4.459 32.108 1509.496 209.652
400 1 400 Large 5679.6 14.191 17.961 32.330 1515.913 842.174
Combined standard uncertainty is
¦
35%
(IEC 60580:2000 / 60601-2-54:2009
and IEC 60601--2--54:2009/AMD1:2015).
Note .