DX- D 600
User Manual
0242F EN 20181218
51
Quantitative Information
Loading Factors Parameter Selection Filtrat. Measured Doses
KVp
mA
Time (s)
mAs
Focal Spot
Selection
SID
Source--Image Distance
(cm)
Collimator blades
opening
(cm)
HVL
(min. value allowed)
(mmAl)
Collimator
Output Dose
(μGy*m2)
SID Dose
(mGy)
Phantom
Input Dose
(mGy)
Phantom
Input Dose Rate
(Gy/h)
Phantom
Outp ut Dose Rate
(mGy/h)
Phantom
Output Dose
(μGy)
160 0.012 2 Small 3.8 0.253 0.395 1 18.573 2493.391 8.311
100 0.1 10 Large
24.4 1.375 2.148 77.331 1679.791 46.661
200 0.5 100 Large
13x13
239.3 14.530 22.704 163.467 3508.591 487.304
400 1 400 Large
882.9 59.548 93.043 133.983 2882.504 2001.739
160 0.012 2 Small
100
38.5 0.262 0.409 122.731 5744.348 19.148
100 0.1 10 Large
210.7 1.444 2.257 81.244 3862.957 107.304
200 0.5 100 Large
43x43
2124.2 15.252 23.832 171.587 8057.739 1119.130
400 1 400 Large
8581.3 62.748 98.043 141.183 6629.009 4603.478
150
160 0.012 2 Small
5.4
3.8 0.077 0.098 29.337 1208.087 4.027
100 0.1 10 Large
24.4 0.426 0.539 19.410 819.235 22.757
200 0.5 100 Large
13x13
239.3 4.548 5.756 41.442 1714.226 238.087
400 1 400 Large
882.9 18.687 23.651 34.057 1409.948 979.130
160 0.012 2 Small
180
38.5 0.080 0.102 30.467 1700.870 5.670
100 0.1 10 Large
210.7 0.453 0.573 20.646 1152.939 32.026
200 0.5 100 Large
43x43
2124.2 4.803 6.078 43.764 2436.730 338.435
400 1 400 Large 8581.3 19.748 24.993 35.990 2005.983 1393.043
Combined standard uncertainty is
¦
35%
(IEC 60580:2000 / 60601-2-54:2009
and IEC 60601--2--54:2009/AMD1:2015).
Note .