Issued: 01-05-2012 Revision: 1.0 Copyright 2012 Agilent Technologies
Page 3 of 13
1. Introduction
1.1 General Information
This standard operating procedure (SOP) defines the procedure for the installation
qualification, operation qualification and performance verification (IQ/OQ/PV) of Agilent 3200
series meters. This procedure provides a guideline to generate the data necessary to validate
or qualify the Agilent 3200 series meters.
This SOP is applicable to all Agilent 3200 series meters that are used in a regulated lab. And
this SOP can be used as a reference for customer to develop their own IQ/OQ/PV documents.
It’s customers’ responsibility to develop and maintain the IQ/OQ/PV SOP for their specific
applications.
1.2 Meter Information
Installation qualification and operation qualification are required for all new meters used in a
regulated environment. The meters will not be released for regulated use until the installation
qualification and operation qualification have been performed with passing result.
Agilent 3200P Benchtop pH Meter
Agilent 3200C Benchtop Conductivity Meter
Agilent 3200D Benchtop Dissolved Oxygen Meter
Agilent 3200I Benchtop Ion Meter
Agilent 3200M Benchtop Multi-Parameter Analyzer
pH, ORP, Conductivity,
Dissolved Oxygen, ISE
1.3 Required Material
The following materials are required to perform the IQ/OQ/PV on Agilent 3200 series meters
Short circuit plug, PN: G4383-40000, shipped with 3200P/3200M/3200I
Conductivity diagnostic tool, PN:5185-8391, shipped with 3200C/3200M
ATC temperature diagnostic tool, PN: 5185-8390, shipped with
3200P/3200M/3200I/3200C/3200D
Manuals/User Guide
For 3200P/3200M with pH function: