ILT.2013
Check W52 for damage, deformation, or excessive RF
path loss.
Is W52 defective?
Go To
RF.3990 Continue
Replace W52 and go to the Troubleshooting Foldout,
block 1 (located just prior to the \Special Cases" tab).
ILT.2015
Remove A6 from the synthesizer and disconnect W54
from A6. Inspect and test W54 for defects
.
Is W54 defective?
Go To
A6.9000 Continue
Replace W54 and go to the Troubleshooting Foldout,
block 1 (located just prior to the \Special Cases" tab).
Agilent 8360 Instrument Level Troubleshooting 1a-25