ILT.2246
Is the synthesizer upper frequency limit
40 GHz?
Go To
ILT.2255 Continue
ILT.2247
Does the RF power problem exist
only
above 26.5 GHz?
Go To
ILT.2255 Continue
ILT.2248
Loose connections or bad RF cabling can easily cause
problems at these frequencies. Check all cabling from
the amp/mult (A28) to the RF OUTPUT (J1). If all RF
cabling appears intact, the most probable cause is the
amp/doubler (A32).
Are the cabling and J1 intact?
Go To
ILT.2260
Go To
RF.3290
1a-38 Instrument Level Troubleshooting Agilent 8360