Location Measurement
Pin 15
0
15 Vdc
Pin 6 +8.0 Vdc
Pin 12 Sweep Ramp (0 to 10 V)
Did all voltages pass?
Go To
ILT.3085 Continue
ILT.3082
Measure at the center pin at each end of the cable
.
Continue
Go To
ILT.3084
ILT.3083
Replace the cable (W23) that connects the rear panel SMI
input to the A10 assembly, and go to the Troubleshooting
Foldout, block 1 (located just prior to the \Special Cases"
tab).
1a-64 Instrument Level Troubleshooting Agilent 8360