instrument and a maximum of 20 meters. Exceeding
this length may lock up the data bus.)
2. The rear panel SMI (A19).
3. The CPU assembly (A15).
4. The ribbon cable W23 (connects A19 to motherboard).
5. There is a remote possibility of a motherboard (A22)
fault.
If these alternatives do not x the problem, consult
\Special Cases" to troubleshoot this problem.
ILT.3100
Check the continuity from the rear panel 10 MHz input to
the end of W16 (connects to A7J1).
Continue
Go To
ILT.3102
ILT.3101
Replace W16 and go to the Troubleshooting Foldout,
block 1 (located just prior to the \Special Cases" tab).
1a-66 Instrument Level Troubleshooting Agilent 8360