Chapter 3 125
Manual Adjustment Procedures
3. DC Log Amplifier Adjustments
3. Press PRESET, CAL, IF ADJ ON OFF (OFF), ADJ CURR IF
STATE.
4. Set the spectrum analyzer controls as follows:
Center frequency .................................................... 15MHz
Span ................................................................................... 0
Resolution bandwidth ........................................... 300kHz
Reference level ...................................................... 10dBm
5. Set up an 3335A as follows:
Frequency ............................................................... 15MHz
Amplitude ............................................................. 10dBm
6. Press MKR, MARKER DELTA on the spectrum analyzer.
7. Decrease the 3335A power to 26 dBm.
8. Calculate the error: EQUATION {\rm Error} = {\rm delta\ marker\
reading}-{\rm 16\ dB} EQUATION.
9. If the error is less than ±0.2 dB, no adjustment is necessary.
10.Set the 3335A power to 10 dBm.
11.Adjust A4R531 (see Figure 3-10, “DC Log Adjustment Locations,)to
read two times the error. For example, if the calculated error is
+0.75 dB, adjust A4R531 for a delta marker amplitude reading of
+1.5 dB. Press
CAL, ADJ CURR IF STATE.
12.Repeat steps 7 through 11.

Table of Contents

Other manuals for Agilent Technologies 8564EC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8564EC and is the answer not in the manual?

Agilent Technologies 8564EC Specifications

General IconGeneral
BrandAgilent Technologies
Model8564EC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals