EasyManuals Logo

Agilent Technologies 8593E User Manual

Agilent Technologies 8593E
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #440 background imageLoading...
Page #440 background image
Chapter 10 441
Analyzer Options and Block Diagrams
Option 190 DVB-C Service Menus
For the 8 MHz channels, R=50, B=38 and A=1 which yields 243.4 MHz.
For the 4 MHz channels, R=35, B=27 and A=0 which yields
246.85 MHz. For the 2 MHz channels, R=49, B=38 and A=2 which
yields 246.571 MHz. The two separate keys,
DECIMAL PLL N and
DECIMAL PLL R, are used to enter the 2 values and contain the default
values for the 8 MHz channel: 9720 (2602 hex) for the N word and
32869 (8065 hex) for the R word.
SHOW AGC VALUES will take a single AGC (signal) reading at the input
of the Option 190 demodulator input and display it both as a hex value
and as its equivalent `gain' (for example, the amount of downconverter
attenuation to subtract). With a signal at the input, by entering a value
into
DNCVRTR ATTEN and reading the AGC value returned by {SHOW
AGC VALUES}}, you can accomplish manually what the DLP automatic
level setting algorithm does. Note that if you want to duplicate the
value of attenuation shown during a measurement and the
measurement set the analyzer attenuator (AT) to 0 dB, you'll also have
to (manually) set AT to 0 dB.
The
AGC HOLD RUN key allows you to freeze the AGC circuit when
toggled to HOLD.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8593E and is the answer not in the manual?

Agilent Technologies 8593E Specifications

General IconGeneral
BrandAgilent Technologies
Model8593E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals