72 Chapter2
Making Adjustments
10. CAL YTF Adjustment Routine
10. CAL YTF Adjustment Routine
This adjustment applies to:
8592L spectrum analyzers
8593E spectrum analyzers
8595E spectrum analyzers
8596E spectrum analyzers
The
CAL YTF softkey accesses an internal adjustment routine. This
routine adjusts the slope and offset of the A3A8 YTF tune voltage for
each harmonic mixing band. The “CAL FREQ Adjustment Routine”
must be performed prior to this adjustment.
Equipment Required
Cable, Type N (m) to SMA (m) (part number 8120-5148)
Additional Equipment for Option 026
Cable, SMA (m) to (m)
Adapter, Type APC-3.5 (f) to APC-3.5 (f)
Additional Equipment for Model 8595E
Cable, BNC, 20 cm (9in)