626 Chapter14
Analyzer Messages
Interpreting Analyzer Messages
CAL: RES BW AMPL FAIL (H)
During the CAL AMPTD routine, the insertion loss of a resolution
bandwidth filter, relative to the amplitude of the 3 kHz resolution
bandwidth filter, was greater than 2.2 dB.
Related Assemblies
A7 analog interface, A11 bandwidth filter, A13 bandwidth filter, A14 log
amplifier/detector.
Troubleshooting Hints
The CAL AMPTD routine will stop at the resolution bandwidth that
failed. If the A11/A13 bandwidth filter assemblies require realignment,
use the crystal and LC bandwidth filter adjustment procedures in
Chapter 2.
If a defective bandwidth filter board is suspected, refer to “IF
Power-Level Measurement” in Chapter 5.
CAL SIGNAL NOT FOUND (U)(H)
During a self-calibration routine, the CAL OUT signal cannot be found.
The primary purpose of the CAL OUT check is to confirm that CAL
OUT is connected to the input. A CAL OUT signal that is not within
300 MHz ±50 MHz, or is not greater than or equal to 45 dBm, causes
this error message to be displayed.
Related Assemblies
A7 analog interface, A9 third converter.
Troubleshooting Hints
If one of the following conditions are present, the error message is
displayed.
The CAL OUT signal is missing. Ensure that the CAL OUT is
connected to the input connector using the CAL OUT cable supplied
with the analyzer.
Refer to Table 1-4 for the part number of the CAL OUT cable.
If the calibration cable is connected properly, manually check the
CAL OUT signal.
The CAL OUT signal is not within ±50 MHz of 300 MHz, or it has an
amplitude less than or equal to 45 dBm. The test limits are large
enough to indicate an obvious failure for either the CAL OUT signal
or the analyzer.
The DEFAULT CAL DATA is in use and the frequency of the CAL
OUT signal appears out of tolerance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8593E and is the answer not in the manual?

Agilent Technologies 8593E Specifications

General IconGeneral
BrandAgilent Technologies
Model8593E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals