EasyManuals Logo
Home>Agilent Technologies>Test Equipment>86100A

Agilent Technologies 86100A User Manual

Agilent Technologies 86100A
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #148 background imageLoading...
Page #148 background image
7-8
Calibration Commands
OUTPut
Initiates a module calibration on a selected module, channel, or slot. For the CHANnel and
SLOT arguments, the specified value should be either 1 (left module position) or 3 (right
module position).
Whenever a calibration message is displayed on the instrument, send the :CALibrate:CON-
Tinue, :CALibrate:CANCel, or :CALibrate:SDONE commands. Sending any other command,
including *OPC, disrupts the instrument forcing you to cycle instrument power.
Example GPIB sequence for module calibration:
10 OUTPUT 707;":CALIBRATE:MODULE:VERTICAL LMODULE" <disconnect all inputs>
20 OUTPUT 707;":CALIBRATE:MODULE:CONTINUE"
30 END
OUTPut
Command :CALibrate:OUTPut <dc_value>
This command sets the dc level of the calibrator signal output through the front-panel CAL
connector.
Example This example puts a dc voltage of 2.0 V on the analyzer Cal connector.
10 OUTPUT 707;":CALIBRATE:OUTPUT 2.0"
<dc_value> dc level value in volts, adjustable from –2.0 V to +2.0 Vdc.
Query :CALibrate:OUTPut?
The query returns the current dc level of the calibrator output.
Returned Format [:CALibrate:OUTPut] <dc_value><NL>
Example This example places the current selection for the dc calibration to be printed in the string
variable, Selection$.
10 DIM Selection$[50] !Dimension variable
20 OUTPUT 707;":CALIBRATE:OUTPUT?"
30 ENTER 707;Selection$
PROBe
Command :CALibrate:PROBe CHANnel<N>
Starts the probe calibration for the selected channel. It has the same action as the command
:CHANnel<N>:PROBe:CALibrate. For more information about probe calibration, refer to
“Probe Calibration” on page 7-3. <N> is an integer, 1 through 4.
Whenever a calibration message is displayed on the instrument, send the :CALibrate:CON-
Tinue, :CALibrate:CANCel, or :CALibrate:SDONE commands. Sending any other command,
including *OPC, disrupts the instrument forcing you to cycle instrument power.
Example 10 OUTPUT 707;":CALibrate:PROBe CHANnel1"
RECommend?
Query :CALibrate:RECommend? {CHANnel<N>}

Other manuals for Agilent Technologies 86100A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 86100A and is the answer not in the manual?

Agilent Technologies 86100A Specifications

General IconGeneral
BrandAgilent Technologies
Model86100A
CategoryTest Equipment
LanguageEnglish

Related product manuals