12-9
Function Commands
VERTical:RANGe
This command defines the full-scale vertical axis of the selected function. This automatically
changes the mode from auto to manual, if the scope is not already in manual mode.
<full_scale_range> is the full-scale vertical range.
Query :FUNCtion<N>:VERTical:RANGe?
This query returns the current range setting of the specified function only when the respec-
tive function display is ON.
Returned Format [:FUNCtion<N>:VERTical:RANGe] <range><NL>
Example 10 OUTPUT 707;":FUNCTION2:DISPLAY ON"
20 OUTPUT 707;":FUNCTION2:VERTICAL:RANGE?"