EasyManua.ls Logo

Agilent Technologies 86100A

Agilent Technologies 86100A
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12-9
Function Commands
VERTical:RANGe
This command defines the full-scale vertical axis of the selected function. This automatically
changes the mode from auto to manual, if the scope is not already in manual mode.
<full_scale_range> is the full-scale vertical range.
Query :FUNCtion<N>:VERTical:RANGe?
This query returns the current range setting of the specified function only when the respec-
tive function display is ON.
Returned Format [:FUNCtion<N>:VERTical:RANGe] <range><NL>
Example 10 OUTPUT 707;":FUNCTION2:DISPLAY ON"
20 OUTPUT 707;":FUNCTION2:VERTICAL:RANGE?"

Other manuals for Agilent Technologies 86100A

Related product manuals