EasyManuals Logo
Home>Agilent Technologies>Test Equipment>86100A

Agilent Technologies 86100A User Manual

Agilent Technologies 86100A
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #315 background imageLoading...
Page #315 background image
21-11
TDR/TDT Commands (Rev. A.05.00 and Below)
STIMulus
ended TDR or TDT measurements. DIFFerential turns on the pulse generator for channels 1
and 2 or channels 3 and 4 for differential TDR or TDT measurements. COMMonmode turn on
the pulse generator for channels 1 and 2 or channels 3 and 4 for common-mode TDR or TDT
measurements. EDIFferential and ECOMmon turn on the pulse generator for channels 1 and
2 (or channels 3 and 4) in either differential or common mode. The pulses are sent to an
external pulse generator and the second pair of channels (3 and 4 or 1 and 2 respectively) are
used as either TDR or TDT destinations.
Restrictions Software revision A.04.20 and A.05.00. TDR mode.
Example The following example turns on pulse generators for channels 3 and 4 for single-ended TDR
measurements.
10 OUTPUT 707;":TDR4:STIMULUS ON3AND4"
Query :TDR{2 | 4}:STIMulus?
The query returns the current settings for the TDR pulse generators.
Returned Format [:TDR{2 | 4}:STIMulus] {OFF | ON | ON1 | ON2 | ON1AND2 | DIFFerential | COMMonmode | EXTernal | ON3 | ON4
| ON3AND4}<NL>

Other manuals for Agilent Technologies 86100A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 86100A and is the answer not in the manual?

Agilent Technologies 86100A Specifications

General IconGeneral
BrandAgilent Technologies
Model86100A
CategoryTest Equipment
LanguageEnglish

Related product manuals