22-8
TDR/TDT Commands (Rev. A.06.00 and Above)
RESPonse:VERTical
where Z
0
equals 50 ohms in the 86100C.
A settings conflict error is reported if no stimulus channel is active. If the response is uncali-
brated, a default value is returned. The value <N> is an integer, 1 through 4, that identifies
the response waveform.
Restrictions Software revision A.06.00 and above. TDR mode.
Example 10 OUTPUT 707;":TDR:RESPONSE1:VAMPlitude?"
Returned Format [:TDR:RESPonse<N>:VAMPlitude] <value><NL>
RESPonse:VERTical
Command :TDR:RESPonse<N>:VERTical {AUTO | MANual}
This command specifies whether the TDR/TDT response should automatically track the
source channel’s vertical scale (AUTO), or use a user-defined scale specified with the VERTi-
cal:OFFSet and VERTical:RANGe commands (MANual). AUTO is the usual setting. The key-
word TSOurce may also be used. This command is compatible with the Agilent 83480/54750
and is equivalent to AUTO. The value <N> is an integer, 1 through 4, that identifies the
response waveform.
Restrictions Available in all software revisions. TDR mode.
Example 10 OUTPUT 707;":TDR:RESPONSE1:VERTICAL MANUAL"
Query :TDR:RESPonse<N>:VERTical?
Returned Format [:TDR:RESPonse<N>:VERTical] {AUTO | MANual}<NL>
RESPonse:VERTical:OFFSet
Command :TDR:RESPonse<N>: VERTical:OFFSet <offset_value>
This command sets the vertical position of the specified response and changes the vertical
tracking setting to MANual if it is in AUTO. Refer to “RESPonse:VERTical” on page 22-8. The
position is always referenced to center screen. The value <N> is an integer, 1 through 4, that
identifies the response waveform. The <offset_value> argument is the offset value in the cur-
rent channel UNITs. Suffix UNITs are ignored; only the scalar part is used (m in mv).
Restrictions Available in all software revisions. TDR mode.
Example 10 OUTPUT 707;":TDR:RESPONSE1:VERTICAL MANUAL"
20 OUTPUT 707;":TDR:RESPONSE1:VERTICAL:OFFSET 50 MV"
Query The information reterned from the query is only valid when the vertical tracking mode is set
to manual.
:TDR:RESPonse<N>:VERTical:OFFSet?
Returned Format [:TDR:RESPonse<N>:VERTical:OFFSet] <volts><NL>
RESPonse:VERTical:RANGe
Command :TDR:RESPonse<N>:VERTical:RANGe <range_value>

Other manuals for Agilent Technologies 86100C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 86100C and is the answer not in the manual?

Agilent Technologies 86100C Specifications

General IconGeneral
BrandAgilent Technologies
Model86100C
CategoryTest Equipment
LanguageEnglish

Related product manuals