1-20
8753ES Specifications and Characteristics
Instrument Specifications
Table 1-19 Test Port Input
8753ES Test Port Input
Dynamic Accuracy (Specification)
For input ports 1 and 2, accuracy of the test port input power reading relative to the reference
input power level.
• Inputs: testport 1 and 2
• For test port powers > −50 dBm and < 0 dBm, magnitude dynamic accuracy is
0.02 dB + 0.001 dB/dB from the reference power, phase dynamic accuracy is
0.132 deg + 0.0066 deg/dB from the reference power.
• For Option 075 and 014, for test port powers up to the maximum source power.